Call for price
The Sykam S 3580 RI Detector is a differential refractive index detector which offers high sensitivity, stability and reproducibility for optimal refractive index detection for HPLC applications.
Description
The Sykam S 3580 RI Detector is a differential refractive index detector which offers high sensitivity, stability and reproducibility for optimal refractive index detection for HPLC applications. The thermally well isolated optics with a countercurrent heat exchanger and programmable temperature control guarantees stable baselines and optimal signal to noise ratio needed for sensitive refractive index detection.
Additional information
| Dimensions | 35 × 22 × 15,5 cm | 
|---|---|
| Detection method | By deflection  | 
		
| Refractive Index range | 1.0 – 1.75  | 
		
| Pressure-flow property | 0.2 – 3.0 ml/min (analytical), 0.2 – 3.0 ml/min (micro), 1.0 – 50.0 ml/min (preparative)  | 
		
| Flowcell volume | 4 µl 45º (micro), 9µl 45º (analytical), 7µl 5º (preparative)  | 
		
| Max flowcell pressure | 6 kg/cm²  | 
		
| Dead volume | 6 µl (micro version), 24 µl (analytical), 88 µl or 353 µl (preparative)  | 
		
| Linearity range | 0 – 500 µRIU (micro version), 0 – 1000 µRIU (analytical), 0 – 20.000 µRIU (preparative)  | 
		
| Temperature setting | Ambient or 35 ºC – 55 ºC by 1ºC steps  | 
		